Rosemount PT100 50 to 450 dg C 0065C21T0135D0
Mfr.
Part Number :- Rosemount
PT100 50 to 450 dg C 0065C21T0135D0
Manufacturer :- Rosemount
Description :- Rosemount PT100 50 to 450 dg C 0065C21T0135D0
Full Description :- The 214C RTD uses the industry-standard PT100 sensor type.
It is available using a thin film element, better in vibration and physical shock, or a wire wound element for a wider temperature range and better for low-temperature applications.
Numerous calibration ranges which include multiple points and Callendar-Van Dusen constants, allow for unmatched accuracy.
Specifications are defined by measurements collected
per methods defined in IEC 60751:2008. Effects of temperature cycling for the
214C RTD are ±0.15°C (0.059Ω) maximum ice-point resistance shift following 10
cycles over the maximum specified temperature range. The hysteresis is ±0.50 °C
(0.185 ) maximum ice-point resistance shift over the maximum specified
temperature range.
Rosemount 214C temperature sensors are available with immersion lengths from 2 to 78-1/2 inches (50 to 2000mm) in 1/4-inch (5mm) increments.
All sensor styles and lengths have a standard 1/4-inch (6mm) nominal diameter.
Multiple options for connection heads, accuracy, mounting styles, extensions and other factors are also available.
The Rosemount 214C Sensors are designed to provide flexible and reliable temperature measurements in process monitoring and control environments. Features include: Temperature ranges of –321 to 1112 °F (–196 to 600 °C) for RTDs and –321 to 2192 °F (–196 to 1200 °C) for thermocouples
Industry-standard
sensor types: PT100 RTDs; thermocouple Type J, Type K, and Type T
Spring-loaded and compact spring-loaded sensor
mounting styles
Hazardous location product approvals and
certification
Calibration services to give insight to sensor
performance
Calibration certificate to accompany sensor
Connection
Head :- Rosemount Aluminum – IP Rating
66/68 –
Conduit /
Cable Entry 1/2-in. NPT
Sensor
lead wire termination :- Terminal block (DIN 43762)
Sensor
type :- RTD, dual element, 3-wire
Temperature
range :- - 50 to 450 °C (- 58 to 842
°F)
Extension
length (N) in millimeters :- 35
mm
Sensor
length (L) in millimeters :- 100mm
IECEx Flameproof Approval
Sensor
calibration:- from - 50 to 450 °C (-58 to 842
°F) with A, B, C, and Callendar-Van Dusen constants
Condition
:- Used /
Working
Price